.. these rolls are defined to use the custom css classes .. role:: yellow .. role:: red .. role:: green .. role:: blue Narda Smarts II Microwave Sensor ================================ The purpose of this procedure is to validate the functionality of the Vault-1 and Hutch-1 Narda Smarts II Microwave Sensor. ----- Starting Conditions ------------------- #. Narda meters are not alarming. #. All meters are calibrated. - Vault-1 - RF-1 #. Vault-1 Control IONIZING RADIATION INTERLOCK protocase MICROWAVE AREA MONITOR lamp is :green:`green`. #. Narda relay contact #10 in Vault-1 Control West aggregator panel LEDs are all on. #. :yellow:`Yellow` and :blue:`blue` contact blocks in Vault-1 Control west aggregator panel have continuity to each other. .. figure:: /images/testing_documentation/narda/Narda.jpg :align: center :scale: 20 % **Figure 1:** This is the Narda Smarts II Microwave Sensors. One is located on the Vault-1 north wall and the other is located on the RF-1 east wall. .. figure:: /images/testing_documentation/narda/yellow_blue_contacts.jpg :align: center :scale: 20 % **Figure 2:** These are the yellow and blue contact blocks in the Vault-1 Control west panel that correspond to the Narda meters. ----- Testing ------- #. Go up to the Narda meter in RF-1and press the red test button. In response: - An audible alarm from the meter. - IONIZING RADIATION INTERLOCK MICROWAVE AREA MONITOR lamp is Vault-1 Control turns :red:`red`. - Relay 10 in the Vault-1 Control west panel shows all diagnostic LEDs on. - :yellow:`Yellow` and :blue:`blue` contact blocks will lose continuity with each other. #. Repeat for Narda meter in Vault-1. .. figure:: /images/testing_documentation/narda/Vault-1_Control_protocase_microwave_fail.jpg :align: center :scale: 20 % **Figure 3:** This is the Vault-1 Control IONIZING RADIATION INTERLOCK protocase MICROWAVE AREA MONITOR lamp when the Narda meter is in alarm. There is no lamp that corresponds to the Hutch-1 Control IONIZING RADIATION INTERLOCK protocase.